YOU98(P)

Technical Parameters

Basic parameters

Model YOU98(P)
Charger input voltage and frequency 220V~50Hz
Power 30VA
Internal power supply 10.8V
Power resistance ≤2Ω
Radiation type X-ray
Ray direction and distribution Beam limiter outlet diameter 60mm
Dose rate 2mGy/s
X-ray tube model Canon D-045
Target Material Tungsten
Anode angle 12.5 °
Ray focus 0.4mm
Tube voltage 70kV ± 10%
Tube current 2mA ± 20%
Exposure time adjustment range 0.04S-2.0S
Nominal electrical power 0.14kW (70 kV, 2mA, 0.1s)
Inherent filtration 1.0mmAl/70kV,YY/T0062-2004
Additional filtration 0.5mmAl/70kV, YY/T0062-2004
Total filtration 1.5mmAL/70kV,YY/T0062-2004
Half price layer >1.6 mmAl at 70kV
Load cycle 1/15
Leakage radiation loading factor exposure for 1 second, stop for 15 seconds
Leakage radiation rate ≤ 0.25mGy/h at 1 meter (70kV, 2mA, 1s, loading interval 1s/15s)
Whole machine weight 1.9kg
Working temperature 10℃-40℃
Storage temperature  -20℃-55℃
Atmospheric pressure 700hPa~1060hPa
Relative humidity at work 30% RH~75% RH
Storage relative humidity ≤93RH
 

Accessories

Videos

F&Q

  • YOU98(P) Maintenance Guide

    YOU98(P) Maintenance Guide


    To ensure optimal performance and long-term use of the X-ray machine, as well as to improve work efficiency and ensure operational safety, the following maintenance instructions must be followed.
    I. Daily Maintenance
    1.1. After using the device, turn off the power and clean/disinfect the surface with 75% alcohol wipes. Do not allow water or other liquids to seep into the interior of the machine to avoid short circuits.
    1.2. During X-ray imaging, patients must wear protective gear, and operators or other personnel should maintain a distance of at least 2 meters from the radiation source of the dental X-ray machine.
    1.3. When the device is not in use, turn off the standby mode to conserve energy.
    1.4. Always operate the device following standard procedures. Do not force shutdown to avoid damaging the battery or circuit board.
    II. Regular Maintenance
    2.1. If the battery is not used for an extended period, recharge it at least once every three months. Only use the original charger provided with the device, and avoid charging for more than 12 hours in a single session.
    III. Precautions for Use
    3.1. To prevent accidental operation, the device is equipped with a lock function (press Mode and Teeth buttons simultaneously to unlock).
    3.2. Do not place the X-ray machine in a location that makes it difficult to quickly power off or disconnect the device.
    3.3. Do not perform maintenance or repair on the device while it is in use with a patient.
    IV. Inspection and Accessory Replacement
    4.1. Check whether the screen displays normally when the device is powered on.
    4.2. Before each use, inspect whether the buttons, buzzer, and indicator lights are functioning properly.
    4.3. Regularly inspect the power cord, plug, and socket to ensure proper connection. Prevent issues such as failure to charge, power short circuit, or electric leakage.
    V. Protection and Storage
    5.1. Prevent electrostatic discharge (ESD) and electromagnetic interference (EMI) from nearby equipment.
    5.2. As the product emits a certain le