DQ-80

Technical Parameters

Basic parameters

Model VRN DQ-80
Control seven-inch touch LCD screen
Handpiece LED Detachable Pieze Handpiece PH-2 ×2,
Sandblasting Handpiece AH-1 ×2 ,AH-2 ×1
Operating Mode G/P/E/A/C
Water Tank 800mL ×1
External Water Hose Unavailable
Power Input 230V~,50Hz
Foot Pedal Wireless Foot Pedal (IPX4) ×1
Wired Foot Pedal (IPX1) ×1
Scaler Tip E1 ×1, P12R ×2, IM1 ×1, P16 ×2, P11 ×2
P12 ×2, P12L ×2, G5 ×1, P1 ×6,G1 ×1
(10pcs in total)
Sterilization Box 2 Sets
Powder Tank Components 2
Small Cart Optional
Weight 6.5kg
Input Voltage 230VAC 50Hz
Input Power 100VA
Vibration amplitude Minimum: 1 um; deviation -50%
Maximum: 100 μ m; deviation +50%
Output half-excursion force Minimum: 0.1 N; deviation -50%.
Maximum: 5 N; deviation +50%
Water pressure 0.01MPa~0.5MPa (0.1bar~5bar)
Compressed air supply 5bar~6bar(0.5 MPa~0.6 MPa)
Frequency 25kHz~35KHz
Tips Hardness 50HRC
Packing Size 310*370*200mm
Storage Environment Temperature: -20℃ ~ 55℃,
Relative humidity: ≤ 90%,
Atmospheric pressure: 70 kPa ~ 106 kPa.
Working Environment a.Ambient temperature: 5℃~40℃
b.Relative humidity: ≤80%
c.Atmospheric pressure: 70kPa~106kPa
d.Temperature of Cooling liquid: 5℃~25℃
Life time 10 year
handpiece can be sterilized 134℃ and 0.22 MPa pressure
 

Videos

F&Q

  • DQ-80 Maintenance Guide

    To ensure the good performance and long-lasting use of the ultrasonic periodontal therapy system, as well as to help improve work efficiency and operational safety, it is necessary to maintain the equipment in accordance with the guidelines.

     

    1. Daily maintenance

    1.1 After tearing off the disposable protective blue film, clean and wipe the surface of the equipment with a damp cloth in time (PS: Alcohol wipes should not be used) to ensure that the equipment is clean, and prevent cross infection.

    1.2 After using the medical water in the endo mode, the pipeline needs to be cleaned with pure water or distilled water. Repeat 1-2 times to prevent the pipeline from being corroded and damaged by the medical water.

    1.3 Before and after sandblasting, the equipment needs to be flushed with the one-click clean function to make sure that the pipeline is smooth and to clear the residual prophylaxis powder in the pipeline.

    1.4 After using the scaler handpiece, clean the surface with water and unscrew the lamp cap at the front end of the handpiece to clean the impurities on the lamp cap and the handpiece lamp position;

    Pull out the sandblasting handpiece (rotate while pulling), rinse the handpiece surface and wipe the tail wire connector position, blow dry the handpiece and the tail wire;

    After cleaning the handpieces, blow dry and sterilize it with 134℃ high temperature and high pressure. It is recommended to use sterilisation bags for sterilisation.

    1.5 Pour out the excess prophylaxis powder in the sand tank after use, and use a medical spray gun to clean the remaining prophylaxis powder in the tank.

     

    2. Regular maintenance

    2.1 The sandblasting handpiece needs to be unclogged regularly. It can be unclogged by blowing through the nozzle with a medical spray gun or by using a sandblasting needle. If the blockage is serious, place the sandblasting handpiece in hot or boiling water for 15 minutes before unclogging.

    2.2 Clean the water tank regularly and wipe the surface to prevent scale accumulation in the water tank.

    2.3 Regularly check and clean the gas/water separator valve at the rear end of the equipment to ensure that the gas source is dry.

     

    3. Inspection and replacement of accessories

    3.1 Before each scaling, please make sure that the mode, power and water output are all adjusted and normal before use.

    3.2 Before each sandblasting, please make sure the air pressure, prophylaxis powder, sand output and water output are all adjusted and normal before use.

    3.3 Regularly check the tail wire connector position, sand tank cover, water tank, air source connector and other accessories to ensure that the sealing rings at the corresponding positions are intact. If worn or damaged, they should be replaced in time to prevent affecting the scaling and sandblasting effects.

    3.4 Before heating the water of the equipment, it is necessary to check and confirm that the water heating switch in the setup is on, so as to ensure that the sandblasting tail wire can produce water normally and prevent dry burning.

    3.5 Before using the wireless foot pedal, please make sure that there are 2 x No.5 batteries in the foot pedal and the wireless foot switch in the setup is on.

     

    4. Protection and storage

    The equipment and consumables should be stored in a dry and ventilated place to avoid moisture and corrosion. In the meantime, the main unit, handpieces and water tank etc., should be protected to avoid impact and damage.