DA-20

Technical Parameters

Basic parameters

Model DA-20
Control Touch Panel
Handpiece LED Detachable Pieze Handpiece PH-1 ×1
Operating Mode G/P/E
Water Tank 800mL ×1
External Water Hose Unavailable
Foot Pedal Wireless Foot Pedal (IPX4) ×1
Wired Foot Pedal (IPX1) ×1
Scaler Tip E14×1,P1×2,G2×1,G1×2,G4×2,E1×1,P4×1,(10pcs in total)
Weight 1.8kg
Input Voltage 230VAC 50Hz
Input Power 35VA
Vibration amplitude Minimum: 1 um; deviation -50%
Maximum: 100 μ m; deviation +50%
Output half-excursion force Minimum: 0.1 N; deviation -50%.
Maximum: 5 N; deviation +50%
Frequency 18kHz~45KHz
Tip output power 3W~20W
Operating mode Continuous running
Tips Hardness 50HRC
Main unit size 285mm*155mm*145mm
Packing Size 475mm*170mm*360mm
Storage Environment Temperature:-20℃~55℃
Relative humidity:≤90%
Atmospheric pressure:70kPa~106kPa
Working Environment Temperature:5℃~40℃
Relative humidity:≤80%
Atmospheric pressure: 70kPa~106kPa
Applicative range of power supply and voltage: 230VAC,50Hz
The maximum inlet pressure of handpiece: 0.5MPa
See the specific specifications on the label.
Life time 10 year
handpiece can be sterilized 134℃ and 0.22 MPa pressure
 
 

Accessories

Videos

F&Q

  • DA-20 Maintenance Guide

    DA-20 Maintenance Guide

    1. Daily Maintenance

    1.1 Before each use, disinfect the handpiece, tips, torque wrench, endodontic wrench, and other accessories to prevent cross-infection.
    1.2 After completing endodontic treatment, replace the medicament with distilled water. Turn the water volume knob to the maximum level and press the foot pedal to flush the tubing for at least 30 seconds; repeat 1–2 times.
    1.3 After treatment, use a wrench to remove the working tip. Wipe the surface of the handpiece and tip with an alcohol swab to remove blood stains. Rinse the handpiece head and connector area with clean water, then rinse the entire handpiece. After rinsing, dry it using a three-way syringe, then place it into a sterilization box for sterilization.
    1.4 After treatment, turn off the power to ensure the main unit is completely powered off. Rinse the surface of the tail cord with clean water, dry it with a three-way syringe, and hang it back on the holder.

     

    2. Periodic Maintenance

    2.1 It is recommended to disassemble and clean the handpiece LED light and light guide column weekly. Unscrew the front part of the handpiece, clean the front shell, light guide column, and LED light in sequence, blow them dry using a three-way syringe, reassemble, and place them in the sterilization box.
    2.2 If the wireless foot pedal will not be used for an extended period, remove the battery.
    2.3 Regularly clean the surface and outlet of the water bottle to prevent scale buildup and blockage. Replace worn-out sealing rings on the water bottle in a timely manner to avoid leakage.

     

    3. Sterilization Instructions

    3.1 The handpiece, working tips, and wrenches should be sterilized at 134°C using high temperature and high pressure. Before sterilization, use compressed air to thoroughly dry any remaining liquid on the handpiece.
    3.2 The working tip must be removed from the handpiece before sterilization, and must not be sterilized together with other instruments.
    3.3 During the sterilization process, always inspect the handpiece for any external damage. Do not apply any protective oil to the surface of the handpiece.
    3.4 When disinfecting with alcohol, soak a piece of clean dry gauze in 70–80% ethanol (by volume), use it to wipe the surface of the part to be disinfected twice, and let it act for 3 minutes. Let it air dry naturally or use a clean, dry cloth to wipe off any remaining disinfectant.

     

    4. Precautions for Use

    4.1 It is generally recommended to use the device for no more than 30 minutes at a time to avoid fatigue.
    4.2 Ensure the handpiece and tail cord socket are dry before connecting.
    4.3 Do not unplug the handpiece while pressing the foot pedal and the device is vibrating.
    4.4 When using the endo rinsing function, set the water volume to maximum and gradually increase the power from level 1 to level 3.
    4.5 When installing endo tips, ensure they are tightly fastened. Do not apply heavy pressure during endodontic rinsing. Do not step on the foot pedal if the endo file is not inserted into the root canal.
    4.6 Refill the water bottle promptly when the liquid level falls below the lower limit to ensure smooth water flow.
    4.7 Do not sterilize the handpiece using the following methods: boiling in solution, soaking in disinfectants such as iodine, alcohol, or glutaraldehyde, or placing it in an oven or microwave for high-temperature baking.

     

    5. Protection and Storage

    5.1 Store the product carefully to avoid impact or vibration. Keep it in a cool, dry, and ventilated place.
    5.2 Do not store together with toxic, corrosive, flammable, or explosive materials.
    5.3 If the product is not used for a long time, it should be powered on and water flowed through it once a month for 5 minutes each time.
    5.4 The product should be stored in an environment with temperatures between –20°C and 55°C, relative humidity of 10%–90%, and atmospheric pressure of 70 kPa–106 kPa.