DQ-80

Технические параметры

Basic parameters

Model VRN DQ-80
Control seven-inch touch LCD screen
Handpiece LED Detachable Pieze Handpiece PH-2 ×2,
Sandblasting Handpiece AH-1 ×2 ,AH-2 ×1
Operating Mode G/P/E/A/C
Water Tank 800mL ×1
External Water Hose Unavailable
Power Input 230V~,50Hz
Foot Pedal Wireless Foot Pedal (IPX4) ×1
Wired Foot Pedal (IPX1) ×1
Scaler Tip E1 ×1, P12R ×2, IM1 ×1, P16 ×2, P11 ×2
P12 ×2, P12L ×2, G5 ×1, P1 ×6,G1 ×1
(10pcs in total)
Sterilization Box 2 Sets
Powder Tank Components 2
Small Cart Optional
Weight 6.5kg
Input Voltage 230VAC 50Hz
Input Power 100VA
Vibration amplitude Minimum: 1 um; deviation -50%
Maximum: 100 μ m; deviation +50%
Output half-excursion force Minimum: 0.1 N; deviation -50%.
Maximum: 5 N; deviation +50%
Water pressure 0.01MPa~0.5MPa (0.1bar~5bar)
Compressed air supply 5bar~6bar(0.5 MPa~0.6 MPa)
Frequency 25kHz~35KHz
Tips Hardness 50HRC
Packing Size 310*370*200mm
Storage Environment Temperature: -20℃ ~ 55℃,
Relative humidity: ≤ 90%,
Atmospheric pressure: 70 kPa ~ 106 kPa.
Working Environment a.Ambient temperature: 5℃~40℃
b.Relative humidity: ≤80%
c.Atmospheric pressure: 70kPa~106kPa
d.Temperature of Cooling liquid: 5℃~25℃
Life time 10 year
handpiece can be sterilized 134℃ and 0.22 MPa pressure
 

Товары

Видео продукции

Загрузить продукт

Вопросы и ответы о продуктах

  • How to do if the tooth cleaning machine does not come out of water?

    If no water flows out when the motor is turned on, try the following steps:

    1.Point the nozzle into the tank, making sure the handle is under the bottom of the tank.

    2.Slide the pause switch to the "On" position (up).

    3.Set the intensity knob to 10.

    4.Hold this position for 30 seconds. If no water flows out after 30 seconds, move to the next step